6855-J033A micro21UV-01 UV/Vis Detector (190 to 600nm)
6855-J043A micro21UV-02 UV/Vis Detector (190 to 900nm)
|
| Wavelength range : |
|
micro21UV-01 : 190 to 600 nm (D2 lamp)
micro21UV-02 : 190 to 900 nm (D2 lamp, W lamp) |
| Spectrum bandwidth : |
|
8 nm |
| Wavelength accuracy : |
|
± 3 nm |
| Wavelength setting reproducibility : |
|
± 0.1 nm |
| Noise level : |
|
± 0.3x10- ABU (250 nm, response : STD, Empty cell) |
| Drift : |
|
± 1.5x10-3 ABU/H
(250nm, Empty cell, constant room temperature) |
| Dimension: |
|
150 (W) x 150 (H) x 500 (D) mm |
| Weight: |
|
Approx. 10 kg |
| Power Input voltage: |
|
AC100 ~ 240 V ±10% 50/60 Hz |
| Power consumption: |
|
145 VA (micro21UV-01), 190 VA (micro21UV-02) |
|
|
|
| Capillary flow cell |
| 50 μm ID Capillary cell (Standard) |
6855-J052A micro21-CPC-50 Capillary UV cell, 50 μm ID
|
| Path length : |
|
8 mm (Effective path length : approx. 1.0 mm) |
| Cell volume : |
|
16 nL |
| Tubing : |
|
50 μm ID x 375 μm OD capillary tube |
| 75 μm ID Capillary cell (Optional) |
6855-J053A micro21-CPC-75 Capillary UV cell, 75 μm ID
|
| Path length : |
|
8 mm (Effective path length : approx. 2.0 mm) |
| Cell volume : |
|
35 nL |
| Tubing : |
|
75 μm ID x 375 μm OD capillary tube |