Do you want to more information?
Please contact us. |
|
|
BP-4340 Automatic Back Pressure Regulator
The BP-4340 is a programmable device for supercritical fluid chromatographs and extraction equipment.
Features:
- Since the dead volume is small (10µL or less), peak components and extracted components do not re-mix giving precise fractioning and separation.
- To control pressure, a valve rod oscillates to repeatedly open and close the valve. The pressure is maintained by changing the time interval between valve opening and closing.
- Since a constant pressure is applied to the extracted components, there are no valve blockages caused by the materials being analyzed.
- The time program enables temperature and pressure changes to be made.
- A flow-through type strain gauge pressure transducer is used to measure the pressure. The unit is designed to minimize dead volume.
- The valve rod inside the valve unit is maintained at a constant temperature by digital PID control.
Specificaions:
Operaing pressure range: |
1.0~50.0MPa |
Operating flow rate range: |
10ml/min max. as liquid carbon |
Pressure control precision : |
+/- 2% of the setting value or +/- 0.2MPa, whichever is larger. |
Temperature setting range : |
35 ~ 80 deg. C (Aluminum block temperature inside the heater jacket) |
|
|